Копия YXQ.FY Series Prefill Syringe Sterilizer

Копия YXQ.FY Series Prefill Syringe Sterilizer

Product Description

Due to the special sealing method of prefill syringes: removable rubber plug at one end, and fixed plug at the other end. Any factor like pressure difference will cause the rubber plugs to move or pop out.YXQ. FY series Prefill Syringe Sterilizer is designed according to the charateristics and sterilizing requirement of prefilled syringes, it conforms to 2010 Version GMP requirement, used for sterilizing and drying of prefill syringes.


Features of Equipment

  • Chamber made of SS 304, jacket made of pressure vessel steel.
  • Special technology for Prefilled Syringes sterilization.
  • Fast cooling process: jacket water-cooling or air-cooling.
  • Fast cleaning process: warm water spraying and cleaning.
  • Fast drying process: hot air circulation.
  • HMEC 2.0 control system, conforms to requirement of GMP.

Parameters

Model Chamber Size
L×W×H
Volume
(m³)
Overall Dimensions
L×W×H
Weight
(kg)
Steam Consumption(kg) Water Consumption Electricity
(kw)
Indutrial steam Pure steam
YXQ.FY-202 1000×500×500 0.25 1260×1360×1900 650 15 5 250 1.3
YXQ.FY-203 1020×550×650 0.36 1280×1380×1900 720 18 7 250 1.3
YXQ.FY-206 1170×610×910 0.6 1430×1440×1900 1450 20 10 300 1.3
YXQ.FY-208 1500×610×910 0.8 1760×1440×1900 1650 22 12 300 2
YXQ.FY-210 1300×680×1180 1 1560×1510×2000 1800 25 15 400 2
YXQ.FY-212 1500×680×1180 1.2 1760×1510×2000 2000 25 15 400 3
YXQ.FY-215 1810×680×1180 1.5 2070×1510×2000 2200 28 17 500 3
Utility steam pressure(0.3~0.5Mpa); water pressure(0.15~0.3Mpa); compress air pressure(0.4~0.7Mpa)